![]() |
Click on the image (when the cursor is a hand) to jump to a topic. |
The Filament control panel displays settings and allows control over the filament (W/LaB6) and high tension settings.
The Filament button switches the heating of the filament on or off. The filament button has three possible settings:
When the filament heating is switched on, the filament is heated
to the value indicated in the Heat to: spin-enter-edit
control at a rate defined by the delay steps.
The Heat to spin
buttons change the filament heating step to which the
filament will be heated. If the filament is off (Filament button is gray), any
changes have no direct effect. If the filament is on (button yellow), the
changes (defined by the spin buttons) are implemented once the Enter
button has been pressed.
Changes in heating current are executed with the delay steps
defined.
The Heat to value cannot exceed the filament limit set.
A progress bar displays the current value of the heating as a fraction of the total range. The total range is determined by the filament limit setting.
The Emission Step spin
buttons change the emission settings of the gun. Press the Enter
button to set the new value. Possible values are in the range 1
to 6. In general, the emission should be set so as not to exceed
normal filament emission values (typically 10 uA for LaB6,
50 uA for W).
At lower high tension values, the emission value must be
increased to keep the emission current similar to the values
obtained at higher high tension settings.
The emission display shows the current value of the emission current.
Pressing the High tension button switches the high tension on and off. The high-tension setting is is the one shown in the drop-down list box on the right. The High tension button has three possible settings:
The high tension is enabled through the High tension enable button on the System On/Off Panel.
The high tension setting is by clicking in the drop-down list box and selecting the required setting (a range of fixed settings, comprising 20, 40, 80, 120, 160 and 200 kV).
![]() |
The status displays (when relevant) waiting time before the filament has reached the status requested by the operator.
Page last modified on 11/26/01