![]() |
Click on the image (when the cursor is a hand) to jump to a topic. |
The Stage Control Panel displays and allows control over specimen-stage positions. The positions stored can be saved to (or loaded) from file using the Stage File control panel in the Stage Control panel flap-out.
The position list displays the specimen-stage positions stored. The list shows columns with X, Y, Z, a and b values (the latter only in case of a double-tilt holder). Positions can be added, deleted, saved in or reloaded fro file, and moved to. Positions are selected by clicking with the mouse (or, if the list has the Windows focus - that is, it is the last Windows item used - the arrow keys of the keyboard can also be used to step through the list).
Instructs the specimen stage to move to the position currently highlighted in the position list.
Specimen-stage Move To's and Go To's
There are two different types of specimen-stage movements
(apart from the ones initiated by the trackball) : Move To's
and Go To's. The difference between these two is in the
procedure followed by the specimen-stage to reach the new
position. With Go To's the specimen stage will move on all axes
(as far as necessary) at the same time. This is the mode of
movement used when the Go To or Undo
buttons are pressed. The Go To is more direct and therefore
faster than the recall.
With Move To's a fixed procedure is used to enhance safety during
movement (to remove the possibility altogether of a high-speed
collision between the moving stage and a fixed part of the
microscope). This fixed procedure consists of the following
steps:
Some of these steps may be absent. For example if a single-tilt holder is used, the b tilt steps are omitted. Or if no new values for X and Y are entered, only a Z go to is done in the third step.
Note: The Move To method is strongly advised if large specimen-stage movements are made.
Instructs the specimen stage to go to the position currently highlighted in the position list.
Specimen-stage Move To's and Go To's
There are two different types of specimen-stage movements
(apart from the ones initiated by the trackball) : Move To's
and Go To's. The difference between these two is in the
procedure followed by the specimen stage to reach the new
position. With Go To's the specimen stage will move on all axes
(as far as necessary) at the same time. This is the mode of
movement used when the Go To or Undo
buttons are pressed. The Go To is more direct and therefore
faster than the recall.
With Move To's a fixed procedure is used to enhance safety during
movement (to remove the possibility altogether of a high-speed
collision between the moving stage and a fixed part of the
microscope). This fixed procedure consists of the following
steps:
Some of these steps may be absent. For example if a single-tilt holder is used, the b tilt steps are omitted. Or if no new values for X and Y are entered, only a Z go to is done in the third step.
Note: The Move To method is strongly advised if large specimen-stage movements are made.
Adds the current specimen-stage position to the position list.
Deletes the currently highlighted position in the position list.
Deletes all positions in the position list. Requires confirmation.
Move To's and Go To's can be done for all five axes together or only for XY. The check box XY only defines which of these methods will be used (XY only on means that only X and Y will be changed). An example of the use of the all-axes method would be tilting to different zone axes on a single crystal, whereas XY only could be used to move to different X-Y positions stored at the currently active tilt settings.
Pressing the arrow button displays the flap-out containing the Control, Set, File and Enable control panels.
Page last modified on 11/13/00