The Stage Control Control Panel provides control over stage settings.
Resets all axes to zero. It is advised to use this function (or at the very least Reset AB before removing a specimen holder from the microscope).
Resets the XY position to 0,0.
Reset the tilts to zero.
The Alpha wobbler tool helps in setting the eucentric height. When the wobbler is switched on (the button will turn yellow), the CompuStage will tilt continuously on a from minus to plus by the value set with the alpha wobbler track bar. By changing the Z height and minimizing the image movement, the CompuStage is set to the eucentric height.
The alpha wobbler track bar sets the alpha wobble angle. The maximum value is 15 degrees (5 degrees for the U-TWIN lens), but the user can set any angle less than that.