Purpose: Aligning all SA images with each
other.
Importance: CONVENIENCE
so that the image remains centered when the magnification is
changed.
Method: Center a recognizable image with the
specimen stage. Lower the magnification one step, center the
image with the Multifunction X,Y knobs. Repeat for all
magnifications.
Procedure
The alignment procedure consists of two or four
steps, dependent on the presence of absence of Parfocal
Magnification Series (last two steps) :
Parfocal magnification series
The parfocal magnification series applies an automatic correction
to each SA magnification so as to make the focus difference
between the SA magnification steps as small as possible. This
correction factor is independent of focus settings and applies
only to the SA magnifications. Similar correction factors for the
M (Mi, and Mh) magnifications as well as diffraction (D) are
already present as standard in the Tecnai software.
Notes: